Specifications:
operating pressure range: -0.05~-1, ..., 0.05~6 bar
Accuracy: 0.02%, 0.05%
pressure port: M20 x 1.5
working fluids: air
|
|
|
|
Tester Model
|
YS-2.5
|
Output Pressure Range
|
bar
|
-0.05~-1
|
0.05~2.5
|
-0.05~-1
|
0.05~6
|
Accuracy
|
%
|
0.02 or 0.05
|
0.05
|
Nominal Area of Piston
|
cm2
|
1
|
Carrier and Piston
|
Nominal Mass
|
kg
|
0.05
|
Produced Pressure
|
bar
|
0.05
|
Supplied Weights
|
Nominal Mass
|
kg
|
0.20.10.05
|
Produced Pressure
|
bar
|
-0.2-0.10.5
|
0.20.10.05
|
-0.2-0.10.5
|
0.20.10.05
|
Quantity Block
|
pcs
|
413
|
322
|
413
|
322
|
Pressure Port
|
M 20×1.5
|
Working Fluids
|
air
|
Overall Dimensions
|
mm
|
590×370×520
|
540×350×520
|
Weight per Unit+Weights
|
kg
|
23
|
27
|